D-M-LOG-HC2-V1_00Rotronic AGBassersdorf, SwitzerlandDocument code UnitLOG-HC2 Universal datalogger:User manualUser manualDocument TypeDocument titlePa
Instructions10OverviewThe logger is a miniaturised universal datalogger for measuring and recording dif-ferent physical measurement parameters. For fu
Instructions11Turning the unit on / saving measurement parametersThe storage properties of the logger are determined using the Setup program*.Control
12InstructionsLogger display / Logger operationWhen the logger is turned off 1 the display cannot be activated using the T1 or T2 keys. The display an
Instructions13Logger display and logger operationMenu 0Record controlState: on/stop by KEYChannel indication 0Logger name,model, version, se-rial nu
Instructions14Indicator LED’s Information displayedLED Meaning Red Alarm limit reached Blue Define the behaviour of the blue LED via Setup > Ba-sic
Instructions15MaintenanceUser checks• Check the logger before each use. • Before using the logger check for visible signs of damage.• Check the functi
Instructions16SpecificationsSee data sheetsDefining the axes-Y-Z-ZY-XX
Instructions17Terminal assignment (only LOG-HC2-RO1)Alarm:•AnalogInput(internal)AIN1...AIN4:RECStart/Stop:GND:•Ground•RecordStart/St
Instructions18Operating, transport and storage conditions• Protect the logger from excessive exposure to the sun and other sources of heat. Avoid hea
Instructions19• LOG-HC2-RO1, or LOG-HC2-P1• CD including: • User manual • Logger PC software• Logger USB connection cable• Mains adapterOptions: •
D-M-LOG-HC2-V1_00Rotronic AGBassersdorf, SwitzerlandDocument code UnitLOG-HC2 Universal datalogger:User manualUser manualDocument TypeDocument titlePa
SetupOnlineViewerReaderPC software20SetupReaderViewerOnline
21SetupReaderViewerOnlineOverviewExternal processing of logger data is carried out using the LOG PC software pro-grams Setup, Reader, Viewer and Onlin
22SetupReaderViewerOnlineVersion number of CD Version numbers of LOG PC programs Logger program windowStarting the logger SoftwareThe logger software
23SetupReaderViewerOnlineUninstallingThe software is uninstalled via the computer’s operating system (Programs > LOG_HC2 > Uninstall LOG_HC2).Sy
24SetupReaderViewerOnlinePreparationBefore using the LOG PC programs Setup, Reader and Online, the following preparations must be completed:• Use the
25SetupReaderViewerOnlineTemplateThe template, selected via Setting > LOG-HC2 (*.mse), defines which sensors the Reader reads out as stand
26SetupReaderViewerOnlineOnce the data has been transferred to the PC vari-ous options are available to the user. Details may be found in the text win
27SetupReaderViewerOnlineCompletion of the preparatory measuresUpon exiting Settings the PC saves the settings that were last entered.Once entry of th
28Setup is used to select the sensors for which the logger is required to save measurement parameters, to enter the measurement freque
29->30SetupBasic settingsThis page describes the most commonly used settings – the Basic settings. Infor-mation on further settings can be found on
D-M-LOG-HC2-V1_00Rotronic AGBassersdorf, SwitzerlandDocument code UnitLOG-HC2 Universal datalogger:User manualUser manualDocument TypeDocument titlePa
30SetupPredictionThe maximum amount of data that can be re-corded is influenced by:• The number of sensors• The measurement rate • The battery capacit
31Reducing the volume of dataWhen making extensive recordings it is recommended that all sensors that are not required be “turned off”. This prevents
32LED behaviourThe behaviour of the blue LED is defined via Setup > Basic settings.Situation Behaviour of the blue LEDStart time has been transferr
33SetupUser settingsTo display the configuration saved in the logger select the Basic settings tab 1 and click Read basic settings. The settings are n
34SetupAdjusting the channel propertiesSelect the tab Channels 3 in order to change the name, unit, display format and calibration of a channel. The f
35SetupUsecalibrationUser informationBasic unitUser information Display unitLogger and PC displayUser informationNameUnitDisplay formatYb = Gb * X +
36SetupChange basic parameterChange basic parameter is used to adapt the factory settings (name, unit, display format) to your requirements. These det
37SetupCalibrating the sensors The sensors and analogue inputs can be calibrated using a 2 point calibration process. Those sensors marked “not
38SetupResetting the calibration to the factory defaultsUncheck the checkbox 4 (Apply calibration) and transfer the change to the logger by clicking W
39SetupDefining the logger displaySelect Display and then click Read user settings at the bottom.Define what the device should display following a pow
D-M-LOG-HC2-V1_00Rotronic AGBassersdorf, SwitzerlandDocument code UnitLOG-HC2 Universal datalogger:User manualUser manualDocument TypeDocument titlePa
40Deletes all measurement parameters saved in the loggerFormatting the memoryCAUTIONFormtting will erase all measurement param-eters saved in the logg
41ReaderWith the Reader users can selectively transfer data records logged with the log-ger to a PC. It does however free measurement records that hav
42ViewerViewerRecords created in Reader or Online may be viewed and edited on a PC with the Viewer. The measurement parameters may be displayed either
43ViewerMove traces With right mouse button held downEnlarge section Mark the required section with the left mouse button held down. See also Graphi
44ViewerFileThe File menu is used to Open records , to Reopen (Open again) the most recently used records and to Save the currently open recor
45ViewerView, Graph ,Table, Configuration GraphGridlines can be shown for each axis (bottom, left, right).Fixed axis simplifies the amount of de
46ViewerTable (Tabelle)Go to time allows users to jump straight to the line in the table with the required time.Jump to beginning of graph causes
47ViewerUsing Configuration >Time axis X, the time axis can be annotated, the time segment to be displayed can be set, and the Increment for the
48OnlineOnline allows users to view the progress of the measurement parameters directly on-screen.Procedure:• Complete the preparations -&g
49OnlineFor recurring applications in Online, Templates assist in se-lecting the required sensors and adjusting the display to suit.Create template: S
User manual
50 CSVConcatCalcCutterUtility programs
51The CSV utility creates text files (*.csv or * .fmc) from data records (*.rog). Files saved in *.csv format can subsequently be opened and edited in
52The Cutter utility creates an extract from a data record (*.rog). Creating an extract from a data record• Start the Cutter utility (Logger Program w
53CalcCalcWith Calc curves from existing data records can be linked to each other using for-mulae and saved as a data record (*.rog). The saved data r
54CalcModify the order of the new curvesWith the left mouse button pressed, drag the number field to the required position.Delete new curvesClick insi
55CalcOperators / brackets+ - * / ^ Plus, minus, multiply, divide, to the power of( ) Open brackets, close bracketsFunctionssqrt(no.) Square roo
56ConcatConcatWith Concat (concatenate = to link) data records can be linked together and saved as a new data record (*.rog). The newly cr
57Concat Delete all data records from the list Use Delete all to remove all data records from the list. Delete one data record from the list Sele
58 FreeMatAdditional software for PCs
59FreeMatFreeMat is open source math software that can be installed from the logger CD (ADDITIONAL\FREEMAT). FreeMat can be used to carry out
Operating instructionsUtility programsAdditional software for PCsPC softwareSetupReaderViewerOnlineSetupReaderViewerOnlineInstructions
11.2012LOG-HC2 Firmware V1.12LOG-HC2 PC-Software V2.0060
Instructions7ViewerOnlineImportant notes regarding this user manualIn this manual notes of particular importance are presented as follows:CAUTIONIndic
Instructions8ViewerOnlineSafety instructions and warningsWARNING• Read the operating instructions carefully before using the logger or the logger soft
InstructionsOperating InstructionsLOG-HC2-RO1LOG-HC2-P1Operating Instructions