Rotronic LOG-HC2 Universal Datalogger Uživatelský manuál Strana 38

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 60
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 37
38
Setup
Resetting the calibration to the factory defaults
Uncheck the checkbox
4
(Apply calibration) and transfer the change to the logger
by clicking Write user settings.
CAUTION
If, following calibration, changes are made via
„Changing the basic parameter“ on page 36, the
performed calibration will be lost.
Changing the display parameter
Changes that are made with Change display parameter have no effect on the
settings made with Change basic parameter. This allows you to always perform
calibration using the units selected with Change basic parameter.
Select the desired channel (e.g. analogue input A2) and then click Change dis-
play parameter. Enter your adjustments
5
(e.g. psi instead of Mbar).
To enter the conversion factor (e.g. Mbar to psi) the following options are available:
Enter the amplification and the zero point
Using the calculation assistant
6
enter the values for points P1 and P2
7
. The
program will then calculate the amplification and the zero point.
Activate “Apply user information” by checking the checkbox
4
and then click
OK. Transfer the changes to the logger by clicking Write user settings.
3
4
6
6
Max. 4 characters
7
5
Resetting the properties to the factory settings:
Uncheck checkbox
4
(Apply user information) and transfer the changes to the
logger by clicking Write user settings
2
.
Zobrazit stránku 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 59 60

Komentáře k této Příručce

Žádné komentáře